• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC FUSION ENT NAVIGATION SYSTEM; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC FUSION ENT NAVIGATION SYSTEM; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9733560XOMR
Device Problem Failure to Shut Off (2939)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/03/2020
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: 9733467, logs were received and analyzed.A software anomaly was confirmed.This anomaly is tracked through the medtronic database.Codes b01, c10 and d02 are applicable.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system that was used outside of procedure.It was reported that while shutting down the system the monitor went to a black screen and the power light was yellow.The system was shutoff, re-booted, and was able to shut down normally.No patient present.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION ENT NAVIGATION SYSTEM
Type of Device
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
tricha miles
7000 central avenue ne rcw215
minneapolis, MN 55432
7635140379
MDR Report Key10983553
MDR Text Key220688900
Report Number1723170-2020-03233
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994594235
UDI-Public00613994594235
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9733560XOMR
Device Catalogue Number9733560XOMR
Device Lot Number4501028573
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2020
Date Manufacturer Received12/01/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/04/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-