Model Number 1012-04-050
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/25/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Original surgery don (b)(6) 2020.Right hip revision.Stem subsided - frequent dislocations.Did the patient experience a post-op device malfunction? --> unknown, did the patient experience an adverse event such as infection, non-union, allergic reaction, osteoporosis, overloading, pain, degenerative diseases, bleeding or oozing? --> unknown, did the patient require revision surgery or hardware removal? --> yes, was device explanted? --> false, did patient require revision surgery? --> false, patient status/ outcome / consequences --> unknown, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: --> unknown, is the patient part of a clinical study --> unknown, (b)(4).Device property of -->none, device in possession of -->none, (b)(4).Device property of -->none, device in possession of -->none, (b)(4).Device property of -->none, device in possession of -->none.By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.--> true.
 
Event or Problem Description
Additional information received stating that the disassociation of the ceramic head and serf liner, as well as subsidence of the femoral stem is what led to dislocation.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
1.Is it a case of dislocation between the ceramic head and serf liner? answer: the disassociation of the ceramic head and serf liner, as well as subsidence of the femoral stem is what led to dislocation.2.Context: traumatic event? fall? answer: there was no traumatic event that was brought to my attention.3.It was mention "frequent dislocations", can you please provide historical data on this.Answer: the patient experienced multiple dislocations.I am not sure of the exact number, or when they occurred.4.Please provide x-ray.Answer: i do not have access to the x-ray.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-LOCK BPS SZ 5 STD OFFSET
Common Device Name
TRILOCK HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10983736
Report Number1818910-2020-26699
Device Sequence Number3235102
Product Code LZO
UDI-Device Identifier10603295000761
UDI-Public10603295000761
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K073570
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1012-04-050
Device Catalogue Number101204050
Device Lot NumberJ68H36
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/25/2020
Supplement Date Received by Manufacturer12/16/2020
12/16/2020
01/19/2021
Initial Report FDA Received Date12/10/2020
Supplement Report FDA Received Date12/28/2020
01/14/2021
01/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BI MENTUM PFRK PE LINER 28 49; BI MENTUM PFRK PFIT CUP 49; DELTA CER HEAD 12/14 28MM +5; BI MENTUM PFRK PE LINER 28 49; DELTA CER HEAD 12/14 28MM +5
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-