BOSTON SCIENTIFIC CORPORATION ANGIOJET ULTRA 5000A; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 45038
Medical Device Problem Code Failure to Disconnect (2541)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the footswitch failed to release and the procedure was cancelled.An angiojet ultra 5000a was selected for a thrombectomy procedure.During preparation, it was noted that the unit was operating like the foot pedal is being depressed even though it is not in use.The procedure was cancelled.No patient complications were reported.
 
Event or Problem Description
It was reported that the footswitch failed to release and the procedure was cancelled.An angiojet ultra 5000a was selected for a thrombectomy procedure.During preparation, it was noted that the unit was operating like the foot pedal is being depressed even though it is not in use.The procedure was cancelled.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the ultra foot switch pedal was received by san jose cis in bad condition with physical damages/defects observed.The unit failed visual inspection and mechanical test due to depressed plate on press step test therefore this issue was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET ULTRA 5000A
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10983770
Report Number2134265-2020-17241
Device Sequence Number10938477
Product Code MCX
UDI-Device Identifier08714729890584
UDI-Public08714729890584
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P980037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/07/2025
Device Model Number45038
Device Catalogue Number45038
Device Lot NumberOEMTRACE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2021
Initial Date Received by Manufacturer 11/30/2020
Supplement Date Received by Manufacturer05/05/2021
Initial Report FDA Received Date12/10/2020
Supplement Report FDA Received Date05/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
92069759-FA
Patient Sequence Number1
-
-