| Model Number |
45038 |
| Medical Device Problem Code |
Failure to Disconnect (2541)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the footswitch failed to release and the procedure was cancelled.An angiojet ultra 5000a was selected for a thrombectomy procedure.During preparation, it was noted that the unit was operating like the foot pedal is being depressed even though it is not in use.The procedure was cancelled.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the footswitch failed to release and the procedure was cancelled.An angiojet ultra 5000a was selected for a thrombectomy procedure.During preparation, it was noted that the unit was operating like the foot pedal is being depressed even though it is not in use.The procedure was cancelled.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the ultra foot switch pedal was received by san jose cis in bad condition with physical damages/defects observed.The unit failed visual inspection and mechanical test due to depressed plate on press step test therefore this issue was confirmed.
|
| |
|