| Model Number |
TRANSDUCER S7-3T |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
|
| |
|
Event or Problem Description
|
|
A customer reported an s7-3t model transducer had an articulation issue during use.There was no injury associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the transducer confirmed the imaging issue was rotational and not an articulation failure.The transducer failed parallel resistance and image rotation performance tests.Investigation of the device revealed dried fluid ingress and corrosion in the cable, a dent in the cable jacket, and scratches to the beading, window, and tipshell.The physical damage to the transducer inhibited the overall performance of the device and is indicative of improper maintenance.
|
| |
|