MAUDE Adverse Event Report: S7-3T; TRANSDUCER, ULTRASONIC, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

S7-3T; TRANSDUCER, ULTRASONIC, DIAGNOSTIC Back to Search Results
Model Number TRANSDUCER S7-3T
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
 
Event or Problem Description
A customer reported an s7-3t model transducer had an articulation issue during use.There was no injury associated with this event.
 
Additional Manufacturer Narrative
Evaluation of the transducer confirmed the imaging issue was rotational and not an articulation failure.The transducer failed parallel resistance and image rotation performance tests.Investigation of the device revealed dried fluid ingress and corrosion in the cable, a dent in the cable jacket, and scratches to the beading, window, and tipshell.The physical damage to the transducer inhibited the overall performance of the device and is indicative of improper maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S7-3T
Common Device Name
TRANSDUCER, ULTRASONIC, DIAGNOSTIC
MDR Report Key10984036
Report Number3019216-2020-00117
Device Sequence Number17208116
Product Code ITX
UDI-Device Identifier00884838061439
UDI-Public00884838061439
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K132304
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTRANSDUCER S7-3T
Device Catalogue Number989605406772
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date03/19/2021
Initial Date Received by Manufacturer 11/27/2020
Supplement Date Received by Manufacturer11/27/2020
Initial Report FDA Received Date12/10/2020
Supplement Report FDA Received Date04/06/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/16/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
-
-