Model Number 54794
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).The customer is located in (b)(6) where the meter part number is 54794.As the manufacturer, a similar device part number for the us to report is 54790.Therefore, this report is being submitted to the fda under 54790 and the subsequent report to health (b)(6) will reflect 54794.The customer reported that there was a barcode misread on a statstrip meter.The facility reported that the meter scanned a patient barcode in the emergency department, however was unsuccessful in obtaining accurate details of the discrepant read.The customer stated that the information was detained in their system and they were unable to determine who the patient was and unable to obtain the armband.The facility is continuing to utilize the meter, therefore; no sample was returned to nova for evaluation.There was no reported patient harm or intervention.A device history record (dhr) review for the analyzer was performed by the investigator.The review included an assessment of the production, testing, and release of the products.No abnormalities or concerns were observed; the dhrs indicated that the released product met all specifications.Previously opened and investigated incidents from the customer facility determined that the root cause for the patient misreads is attributed to the user facility using poor quality 1d barcodes that were printed using code -39 symbology.Code-39 is no longer industry standard, and has a higher substitution failure rate than code-128 and much higher than the standard 2d barcodes available.This failure mode does not represent a systemic issue or a failure of the meter to perform as intended.The failure mode is directly related to the quality of the barcode being generated by the customer.The root cause of the misread is most likely due to poor barcode quality.Based on the original investigation, nova suggested to prevent barcode mis-reads in the future, that the customer should change from a 1d to a 2d barcode, which would greatly reduce the chances of a barcode misread.Trending will be monitored for this and or similar complaints.
 
Event or Problem Description
The customer reported that there was a barcode misread on a statstrip meter.The facility reported that the meter scanned a patient barcode in the emergency department, however was unsuccessful in obtaining accurate details of the discrepant read.The customer stated that the information was detained in their system and they were unable to determine who the patient was and unable to obtain the armband.The facility is continuing to utilize the meter, therefore; no sample was returned to nova for evaluation.There was no reported patient harm or intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STATSTRIP GLUCOSE HOSPITAL METER SYSTEM
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
NOVA BIOMEDICAL CORP
200 prospect street
waltham MA 02454 9141
MDR Report Key10985011
Report Number1219029-2020-00052
Device Sequence Number3813119
Product Code PZI
UDI-Device Identifier00385480547901
UDI-Public00385480547901
Combination Product (Y/N)Y
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number54794
Device Catalogue Number54790
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2020
Initial Report FDA Received Date12/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-