• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EMPOWR HIP; EMPOWR ACETABULAR SYSTEM, LINER, 10° HOODED, HXE+, 36G

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EMPOWR HIP; EMPOWR ACETABULAR SYSTEM, LINER, 10° HOODED, HXE+, 36G Back to Search Results
Model Number 942-01-36G
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 11/12/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - patient was treated for possible infected joint.
 
Additional Manufacturer Narrative
Corrected data: see g.1.Manufacturer narrative: the reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 34 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post-surgical instructions.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR HIP
Common Device Name
EMPOWR ACETABULAR SYSTEM, LINER, 10° HOODED, HXE+, 36G
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key10985076
Report Number1644408-2020-01161
Device Sequence Number3299205
Product Code LPH
UDI-Device Identifier00190446309772
UDI-Public(01)00190446309772
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K190057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/09/2025
Device Model Number942-01-36G
Device Catalogue Number942-01-36G
Device Lot Number623Z1015
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/12/2020
Supplement Date Received by Manufacturer02/12/2021
Initial Report FDA Received Date12/10/2020
Supplement Report FDA Received Date03/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-362 LOT 866B1704; 425-97-005 LOT 170R1035; 940-00-030 LOT 147Z1016; 940-00-035 LOT 148Z1027; 940-02-54G LOT 801Z1037; 400-03-362 LOT 866B1704; 425-97-005 LOT 170R1035; 940-00-030 LOT 147Z1016; 940-00-035 LOT 148Z1027; 940-02-54G LOT 801Z1037
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-