Model Number RBYPCLP10
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2020-02238.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the lumbar artery using packing coil lps and a non-penumbra microcatheter.During the procedure, a packing coil lp would not advance past its initial position within its introducer sheath; therefore, it was removed.Next, the same issue occurred with another packing coil lp.Therefore, this packing coil lp was also removed.The procedure was completed using additional lp coils and the same microcatheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the first returned packing coil lp revealed that the pusher assembly mid-joint was retracted through the introducer sheath friction lock.If this occurs, resistance may be experienced during advancement.Forceful advancement against this resistance likely contributed to the kinks observed near the pusher assembly mid-joint.During functional testing, resistance was experienced while attempting to advance the packing coil lp from its returned position within its introducer sheath and, consequently, the pusher assembly kinked further.Evaluation of the second returned packing coil lp revealed that the pusher assembly mid-joint was retracted through the introducer sheath friction lock.If this occurs, resistance may be experienced during advancement.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance.The packing coil lp was then advanced through its introducer sheath without an issue.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 1.3005168196-2020-02238 h3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKING COIL LP
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10985152
Report Number3005168196-2020-02237
Device Sequence Number14196983
Product Code HCG
UDI-Device Identifier00815948021914
UDI-Public00815948021914
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K192955
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/16/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2025
Device Model NumberRBYPCLP10
Device Catalogue NumberRBYPCLP10
Device Lot NumberF96645
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2020
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/16/2020
Supplement Date Received by Manufacturer01/29/2021
Initial Report FDA Received Date12/10/2020
Supplement Report FDA Received Date02/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age86 YR
Patient Weight77
Date Report Sent to FDA01/01/2005
-
-