• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION REVACLEAR 300; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION REVACLEAR 300; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114748
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 11/11/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a particulate matter was observed inside the cap of a revaclear 300.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information added to h3, h6 and h10 h10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection of the two provided photos showed a wet product out of the original packaging.Photo one showed the product label and photo two revealed the header area had a black particle.The reported condition was verified.The cause could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVACLEAR 300
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10986164
MDR Text Key220970547
Report Number9611369-2020-00183
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414114947
UDI-Public(01)07332414114947
Combination Product (y/n)Y
PMA/PMN Number
K130039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Catalogue Number114748
Device Lot Number0-9621-H-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/17/2020
Initial Date FDA Received12/10/2020
Supplement Dates Manufacturer Received01/08/2021
Supplement Dates FDA Received01/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-