| Model Number |
10617 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that stent fracture occurred.A 4.50 x 12 synergy ii drug-eluting stent was advanced for treatment.However, the stent strut were fractured and broke into two pieces.The procedure was completed with a second stent.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: synergy ii us mr 4.50 x 12 mm was returned for analysis without the delivery system.A visual examination of the returned stent found that it was dislodged from the delivery system and damaged on its entire length with strut rows expanded and stretched.An examination under microscope of the stent could not find evidence of any broken struts/connectors.The balloon could not be reviewed as it was not returned.The tip of the device could not be reviewed as it was not returned.The hypotube shaft could not be reviewed as it was not returned.The shaft polymer extrusion could not be reviewed as it was not returned.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that stent fracture occurred.A 4.50 x 12 synergy ii drug-eluting stent was advanced for treatment.However, the stent strut were fractured and broke into two pieces.The procedure was completed with a second stent.There were no patient complications nor injuries reported.
|
| |
|