| Model Number |
6947M72 |
| Medical Device Problem Codes |
Over-Sensing (1438); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead had a lead integrity alert (lia) triggered by high rate non-sustained episodes and sensing integrity counter (sic).In addition, the lead had high thresholds and t-wave oversensing (twos) and oversensing was noted on stored electrograms (egm).The lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|