Model Number CD3361-40C
Device Problem Failure to Deliver Shock/Stimulation (1133)
Patient Problems Syncope (1610); Arrhythmia (1721)
Event Date 11/27/2020
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
During an in clinic follow-up, the device was being interrogated and many stored episodes were being downloaded.The patient experienced ventricular tachycardia (vt) and the physician attempted to provide non-invasive pacing stimulation, however, this was unavailable due to the loading episodes.The patient began experiencing syncope and the physician had to deliver external defibrillation to resolve vt and syncope.The loading of the episodes completed and the event was resolved.The patient was in stable condition following the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ABBOTT
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key10987952
MDR Text Key220884951
Report Number2938836-2020-09561
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Model NumberCD3361-40C
Device Lot NumberA000029320
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/27/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-