• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9438-05
Medical Device Problem Code No Device Output (1435)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that transmitter failed error occurred.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).B5: describe event or problem - additional.D9: device availability ¿ additional.G3: date received by manufacturer - additional.H2: additional information/device evaluation.H3: device evaluated by manufacturer - additional.H6: event problem and evaluation codes - additional.
 
Event or Problem Description
It was reported that transmitter failed error occurred.The product was evaluated.An external visual inspection was performed and passed.Voltage check was performed and passed.Pairing test/download was performed and passed.A review of the share logs was performed and transmitter failed error was not found within the investigation window.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
MDR Report Key10988486
Report Number3004753838-2020-153884
Device Sequence Number19020742
Product Code MDS
UDI-Device Identifier00386270000125
UDI-Public00386270000125
Combination Product (Y/N)N
Initial Reporter CountryEZ
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9438-05
Device Catalogue NumberSTT-GL-004
Device Lot Number5271284
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/18/2020
Supplement Date Received by Manufacturer03/22/2021
Initial Report FDA Received Date12/10/2020
Supplement Report FDA Received Date03/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-