| Model Number |
7206011 |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the 4.5 mm full radius bonecutter was rusted.This was noticed during set up of the shoulder arthroscopy procedure; therefore, the device was not used in the patient.A backup device was available and no significant delay was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).The reported device was received for evaluation.A visual inspection revealed the device was not returned with original packaging and the does have some discoloration looks to have been from heat.No indications of rust no physical damage is visible to the device.A functional evaluation of the returned device found that device does work.Device was hooked up and powered up.Device passed all functional test, a complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate a review of the inspection process found that each component should be visually inspected for all non-conforming attributes defined in the procedure.It was determined the device contributed to the reported event.The complaint was confirmed, and the root cause was associated with component failure.Factors that could have contributed to the reported event include inadequate irrigation during use, or an impact event inconsistent with normal use.Periodic irrigation of the blade is recommended to provide adequate cooling of the blade and to prevent accumulation of excised materials in the surgical site.Ensure that suction of 128 mmhg minimum is flowing while the instrument is running.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|