SOFRADIM PRODUCTION SAS DEXTILE; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number DXT1612AR
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a robotic transabdominal preperitoneal inguinal hernia repair procedure, the surgeon was unable to insert the mesh on the trocar for it was damaged.Another mesh was used to resolve the issue and complete the procedure.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the product was used outside instructions for use (ifu) specifications.The sample was returned in its original packaging and was found slightly contaminated by blood.The textile knitting pattern and the mesh dimension were found as expected.Almost all of the edges are damaged, cracked.It was reported that the mesh tore.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: it is recommended to use a trocar of at least 10 mm internal diameter to introduce meshes of size 13 x 09 cm and 15 x 10 cm, and a trocar of at least 11 mm internal diameter to introduce a mesh of size 16 x 12 cm.Mesh insertion capability may vary depending on mesh insertion method, mesh size and graspers/trocars used medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXTILE
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
MDR Report Key10991599
Report Number9615742-2020-02828
Device Sequence Number19702623
Product Code FTL
UDI-Device Identifier10884521726536
UDI-Public10884521726536
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K192443
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Model NumberDXT1612AR
Device Catalogue NumberDXT1612AR
Device Lot NumberSUG0682X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2021
Initial Date Received by Manufacturer 11/19/2020
Supplement Date Received by Manufacturer03/12/2021
Initial Report FDA Received Date12/11/2020
Supplement Report FDA Received Date04/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-