| Model Number |
C315HIS02 |
| Medical Device Problem Codes |
Accessory Incompatible (1004); Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during the implant procedure, the catheter and two his bundle leads were broken.It was noted that during the first insertion of the sheath and the his lead, the physician tried locating a target spot, but there was no good his signal, therefore the physician decided to change the location.The physician removed the lead to inspect the helix and reinserted back into the catheter sheath and found some resistance during insertion.The physician removed the sheath and lead and found a ¿cut like tear on the sheath¿.The catheter was replaced.The physician attempted to screw in the his lead with much resistance.The physician turned the lead multiple times, but the lead ¿did not progress¿.The physician removed the lead and noticed that the lead was broken at an angle.The lead was replaced, and a similar ¿fate¿ occurred with another his bundle lead.The his bundle lead was not used, and another lead was implanted.The newly implanted lead had high initial impedance and threshold measurements but dropped during closing of the pocket.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the delivery catheter was returned and analyzed.The mechanical operation of the catheter shaft was kinked/buckled at 22 cm from the proximal end.Visual analysis of the catheter indicated damage during use.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|