• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SLOTTED MALLET; HIP INSTRUMENTS : IMPACTION INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SLOTTED MALLET; HIP INSTRUMENTS : IMPACTION INSTRUMENTS Back to Search Results
Model Number 6020-17-000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The primary surgery was performed on (b)(6) 2020 via tka.It was reported that during the surgery, the white plastic part of the mallet came off.There was no harms to the patient.No further information is available.
 
Additional Manufacturer Narrative
Product complaint # : pc-(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the head component does not firmly thread onto the mallet, however, no breakage or missing material was identified.The need for corrective action is not indicated.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLOTTED MALLET
Common Device Name
HIP INSTRUMENTS : IMPACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10992224
Report Number1818910-2020-26809
Device Sequence Number19027386
Product Code GFJ
UDI-Device Identifier10603295178439
UDI-Public10603295178439
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6020-17-000
Device Catalogue Number602017000
Device Lot NumberAH037061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2020
Initial Date Received by Manufacturer 11/25/2020
Supplement Date Received by Manufacturer12/30/2020
01/29/2021
Initial Report FDA Received Date12/11/2020
Supplement Report FDA Received Date01/12/2021
02/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-