Model Number M00558620
Medical Device Problem Codes Leak/Splash (1354); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Investigation result a visual examination of the returned complaint device found that the balloon and catheter did not have any visual defects.Functional analysis was performed, and the balloon was inflated without a problem; however, the balloon would not hold pressure due to several pinholes in the balloon.The pinhole confirms the reported event of the balloon leaking.This failure is likely due to factors or conditions related to the procedure during the use of the device, such as the interaction with the scope or any other surface in the papilla anatomy, that created friction that could have caused the balloon pinhole.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre pro wireguided dilatation balloon was used in the papilla during an endoscopic papillary balloon dilation (epbd) and endoscopic balloon sphincteroplasty procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the balloon was injected with contrast; however, it was noticed that the balloon was leaking.The procedure was completed with another cre pro wireguided dilatation balloon.There were no patient complications reported as a result of this event.Investigation results revealed that the balloon had a pinhole; therefore, this is now an mdr reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE PRO
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10992489
Report Number3005099803-2020-05936
Device Sequence Number14377381
Product Code FGE
UDI-Device Identifier08714729797500
UDI-Public08714729797500
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K112994
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/30/2023
Device Model NumberM00558620
Device Catalogue Number5862
Device Lot Number0025482003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/16/2020
Initial Report FDA Received Date12/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age89 YR
-
-