| Model Number |
31MECJ-502 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Codes |
Occlusion (1984); Low Cardiac Output (2501)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, a 31mm masters series valve expanded cuff was selected for implant in the patient.Prior to the procedure the leaflets were checked multiple times using a q-tip, but once the device was implanted one leaflet would not open.The device was removed from the patient and the leaflets appeared to open and close normally outside the annulus.A 33mm medtronic hancock tissue mitral valve was implanted instead and the procedure was completed.The patient did experience a low output syndrome from the 3 aortic cross clamps during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of one leaflet not opening could not be confirmed.No anomalies were found with the valve, and functional testing upon return to abbott indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
|
| |
|