Catalog Number 0165L16
Device Problem Decrease in Pressure (1490)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that water leaked from the balloon of the foley catheter during a pretest.Per follow up via regional partner on 25 nov 2020, it was unknown there was any damage found of the foley catheter.
 
Manufacturer Narrative
The reported event was unconfirmed because the devices meet the specification.The product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.A dhr review was not required as the investigation was unconfirmed.The instructions for use were found adequate and state the following: "1.Method for use: (1) do not inflate the balloon in the urethra.(the urethra may be injured).(2) do not pull the catheter hard.(the bladder/urethra maybe injured).2.Applicable patients: (1) patients with delirium who might pull out catheter (when patient tugs at catheter unconsciously, the bladder and urethra may be damaged).Contraindications: 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (4) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients: (1) do not use on patients who are or have been allergic to natural rubber latex." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was evaluated.
 
Event Description
It was reported that water leaked from the balloon of the foley catheter during a pretest.Per follow up via regional partner on 25nov2020, it was unknown whether there was any damage found of the foley catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX LUBRICATH FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10992812
MDR Text Key220959936
Report Number1018233-2020-21597
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 04/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0165L16
Device Lot NumberMYEQ0139
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2020
Initial Date Manufacturer Received 11/18/2020
Initial Date FDA Received12/11/2020
Supplement Dates Manufacturer Received04/05/2021
Supplement Dates FDA Received04/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-