It was reported that while attempting to open the device, it would not open correctly in the distal section of the pipeline.It was stated it looked like a tulip.More than 50% of the pipeline had been deployed when it failed.It was resheathed less than or equal to 2 times.The pipeline was positioned in a distal part of a bend.No additional steps or other devices were required to open the pipeline.The pipeline was resheathed and removed with the microcatheter from the patient.Angiographic result post procedure was good.The patient was undergoing surgery to treat an unruptured, saccular aneurysm located at the opthalmic artery with a max diameter of 5mm and a neck diameter of 4mm.It was noted the vessel tortuosity was severe.There were no patient symptoms associated with the event.The devices were prepared as indicated in the ifu.
|