Model Number 87-5956
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 11/22/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical adverse event received for dislocation - initial.Event is serious and is considered moderate.Event is possibly related to both device and procedure.Treatment: "resume precautions" date of implant: (b)(6) 2020, date of event: (b)(6) 2020, (right hip).
 
Event or Problem Description
Procedure notes (b)(6) 2020 indicate the patient received a right total hip closed manipulation and reduction under anesthesia due to dislocation.The procedure was completed without indication of complication by the surgeon.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Provided x-ray images have been reviewed.A dislocation event is not depicted in any of the images.No evidence of a product problem is identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: g1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S-ROM HEAD FEMORAL COCR 32+9
Common Device Name
S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key10993485
Report Number1818910-2020-26818
Device Sequence Number702223
Product Code JDI
UDI-Device Identifier10603295218869
UDI-Public10603295218869
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K920317
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number87-5956
Device Catalogue Number875956
Device Lot Number8731264
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/30/2020
Supplement Date Received by Manufacturer12/14/2020
01/11/2021
Initial Report FDA Received Date12/11/2020
Supplement Report FDA Received Date12/17/2020
01/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX +4 NEUT 32IDX48OD; PINN CAN BONE SCREW 6.5MMX20MM; PINNACLE SECTOR II CUP 48MM; S-ROM HEAD FEMORAL COCR 32+9; S-ROM*SLEEVE PRX ZTT, 14B-SML; SROM*STM ST,30+4L NK,14X09X130; ALTRX +4 NEUT 32IDX48OD; PINN CAN BONE SCREW 6.5MMX20MM; PINNACLE SECTOR II CUP 48MM; S-ROM HEAD FEMORAL COCR 32+9; S-ROM*SLEEVE PRX ZTT, 14B-SML; SROM*STM ST,30+4L NK,14X09X130
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
Patient Weight68
-
-