• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 3FR SL PROVENA MIDLINE; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 3FR SL PROVENA MIDLINE; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number S4153108D
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of reen1144 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported a midline was placed by an imaging services rn and the infusion center found this line to be leaking.The patient was sent to a different facility for evaluation and replacement as needed.Following assessment of the line, it was determined that a new midline was needed and placed by an imaging services rn.No patient harm was reported.
 
Event or Problem Description
It was reported a midline was placed by an imaging services rn and the infusion center found this line to be leaking.The patient was sent to a different facility for evaluation and replacement as needed.Following assessment of the line, it was determined that a new midline was needed and placed by an imaging services rn.No patient harm was reported.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was unconfirmed because the problem could not be reproduced.The product returned for evaluation was one 3fr s/l provena powermidline midline catheter.Usage residues were observed throughout the sample.Microscopic inspection of the sample did not reveal any damage or evidence of leakage.An attempt to infuse water through the sample using a 12ml syringe revealed the sample to be patent to infusion and aspiration with no observed leaks.No leaks were observed during sustained (>15 seconds) hydraulic pressurization.No deficiencies were discovered during evaluation of the returned sample.Consequently this complaint is unconfirmed at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3FR SL PROVENA MIDLINE
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key10995378
Report Number3006260740-2020-20885
Device Sequence Number14551979
Product Code PND
UDI-Device Identifier00801741153952
UDI-Public(01)00801741153952
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K153393
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberS4153108D
Device Catalogue NumberS4153108D
Device Lot NumberREEN1144
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2020
Initial Date Received by Manufacturer 11/17/2020
Supplement Date Received by Manufacturer12/29/2020
Initial Report FDA Received Date12/12/2020
Supplement Report FDA Received Date01/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient Weight118
-
-