| Model Number |
554-1003-001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a procedure with the tenex system, a portion of the microtip needle separated from the rest of the hand piece.The procedure was completed.There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and evaluated.It was confirmed that the microtip needle had separated from the rest of the handpiece.The fracture site did not immediately indicate a specific cause for the failure.Testing and disassembly of the device did not reveal any further anomalies or defects.
|
| |
|