| Model Number |
404675 |
| Medical Device Problem Codes |
Difficult or Delayed Positioning (1157); Inflation Problem (1310)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
| Date of Event |
07/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
Bioflow vii study: during the introduction of a guiding catheter resistance was felt and the control angio showed a recurrent radial artery perforation.A balloon was inserted to assist and eventually the catheter was withdrawn into the subclavian artery to tamponade the recurrent radial artery, which was successful.As the patient remained asymptomatic it was decided to place the orsiro stent.A underexpansion was noted after implantation and the stent was post dilated a second time.Ivus was performed and demonstrated persistent mild underexpansion.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Neither the complaint instrument nor the angiographic material was returned for analysis.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections during which a pressure test is performed (100 percent control).Based on the conducted investigations, no manufacturing or material related root cause could be identified.
|
| |
|