• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA 5X50 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA 5X50 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1896-5050S
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Fall (1848); Failure of Implant (1924); Implant Pain (4561)
Date of Event 11/18/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.
 
Event or Problem Description
The customer reported that the gamma 3 femoral nail (right) broke at proximal end in-situ, just above lag screw hole.The nail has subsequently sunk and caused the distal locking screws to snap.This nail has been successfully extracted and revised.
 
Additional Manufacturer Narrative
Please note corrections to b1, h6 (clinical signs code and conclusion code).
 
Event or Problem Description
The customer reported that the gamma 3 femoral nail (right) broke at proximal end in-situ, just above lag screw hole.The nail has subsequently sunk and caused the distal locking screws to snap.This nail has been successfully extracted and revised.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.The instructions for use instructs user that: ¿the surgeon must warn the patient that the device cannot and does not replicate a normal healthy bone, that the device can break or become damaged as a result of strenuous activity, trauma, mal-union or non-union and that the device has a finite expected service life and may need to be removed at some time in the future.Explain the need to report unusual changes in the implantation area as well as falls or accidents even if the device or the site of operation did not appear to be harmed at the time.Explain also the need to appear for the postoperative examinations (e.G.X-ray checks) and for the possible explantation of the implant.¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.However, one of the possible causes of the failure is stumbling of the patient over a curb.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
The customer reported that the gamma 3 femoral nail (right) broke at proximal end in-situ, just above lag screw hole.The nail has subsequently sunk and caused the distal locking screws to snap.This nail has been successfully extracted and revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, FULLY THREADED T2 TIBIA 5X50 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
MDR Report Key10996720
Report Number0009610622-2020-00917
Device Sequence Number690713
Product Code HSB
UDI-Device Identifier04546540202512
UDI-Public04546540202512
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K032244
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/31/2024
Device Model Number1896-5050S
Device Catalogue Number18965050S
Device Lot NumberK0623FB
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/18/2020
Supplement Date Received by Manufacturer01/20/2021
02/18/2021
Initial Report FDA Received Date12/12/2020
Supplement Report FDA Received Date02/17/2021
03/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-