• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION TRAUMAONE SYSTEM 2.0X11MM SELF-DRILLING IMF SCREW; SCREW, FIXATION, INTRAOSSEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION TRAUMAONE SYSTEM 2.0X11MM SELF-DRILLING IMF SCREW; SCREW, FIXATION, INTRAOSSEOUS Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported a screw fractured and the screw fragment was left inside the patient.No additional patient consequences have been reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Event or Problem Description
It was reported a screw fractured and the screw fragment was left inside the patient¿s mandible.No x-rays or treatment were required.It was reported that no further information is available.
 
Additional Manufacturer Narrative
Quantity of one 2.0 x 11mm sd imf hex screw was returned for evaluation.Visual examination identified the device to be fractured.However, not all fractured pieces were returned.A fragment of the screw was left in the patient with no intervention/attempt to remove.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAUMAONE SYSTEM 2.0X11MM SELF-DRILLING IMF SCREW
Common Device Name
SCREW, FIXATION, INTRAOSSEOUS
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key10999085
Report Number0001032347-2020-00609
Device Sequence Number3617741
Product Code DZL
UDI-Device Identifier00841036128104
UDI-Public00841036128104
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K983728
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number91-5611
Device Lot NumberJ190580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/18/2020
Supplement Date Received by Manufacturer12/28/2020
09/27/2021
Initial Report FDA Received Date12/13/2020
Supplement Report FDA Received Date01/20/2021
10/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-