Model Number 3186
Device Problems Disconnection (1171); High impedance (1291)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/23/2020
Event Type  Injury  
Manufacturer Narrative
¿date of event¿ is estimated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer reference numbers: 3006705815-2020-33109, 3006705815-2020-33111.It was reported that diagnostics revealed high impedance on the leads.Surgery occurred on (b)(6) 2020 to address the issue.During the surgery, it was discovered that the leads were not fully seated in the ipg.As a result, the leads were plugged back into and fully seated in the ipg to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
PERCUTANEOUS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key10999480
MDR Text Key221230173
Report Number3006705815-2020-33110
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017246
UDI-Public05415067017246
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/15/2022
Device Model Number3186
Device Catalogue Number3186
Device Lot NumberA000101199
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PERCUTANEOUS LEAD.; SCS IPG.
Patient Outcome(s) Other;
Patient Weight57
-
-