| Model Number |
IFNT485 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Edema (1820); Inflammation (1932); No Code Available (3191)
|
| Date of Event |
06/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet (b)(4).Weight unknown / not provided.Fax number unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced peri-implantitis, loss integration, inflammation around the implant and edema.The implant was removed.Tooth #20.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay corrected data and additional information.Associated lot #2012060620 was manufactured prior to 24-sep-2015, as a result, a udi number is not required.The following sections are being reported: b4: date of this report was updated.D4: udi number was removed.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H6: component code was added: 4755.H10: narrative/data was updated.
|
| |
|
Event or Problem Description
|
|
No additional event information at the time of this report.
|
| |
|
Event or Problem Description
|
|
No additional event information at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay additional information.The following sections are being reported: b4: date of this report was updated.D9: device availability was added.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H6: investigation code was added: 3331.H10: narrative/data was updated.Dhr review was completed for the subject lot number (2012060620).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Sterilization record (op#0210) was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number (2012060620) for similar events and no other complaint was identified.As documented in the summary investigations, contributing factors for the reported event likely exist outside of zimmer biomet control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported events remain non-verifiable as they are a medical condition.
|
| |
|
Search Alerts/Recalls
|