MAUDE Adverse Event Report: IMAGER II ANGIOGRAPHIC CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMAGER II ANGIOGRAPHIC CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 38270
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/12/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that tip detachment occurred.A 5fr imager ii angiographic catheter was selected for a splenic artery embolization procedure.A femoral approach was used to access the mildly calcified eccentrically shaped lesion.The anatomy was not tortuous.During the procedure, the tip of the catheter detached inside the patient.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Event or Problem Description
It was reported that tip detachment occurred.A 5fr imager ii angiographic catheter was selected for a splenic artery embolization procedure.A femoral approach was used to access the mildly calcified eccentrically shaped lesion.The anatomy was not tortuous.During the procedure, the tip of the catheter detached inside the patient.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.It was further reported that the imager was being introduced through a 6f sheath at the time of the tip detachment.Approximately 3-5cm of the tip detached.The tip was retrieved with a snare.There were no patient complications.
 
Additional Manufacturer Narrative
Updated: adverse event/product problem, outcomes attributed to adv event, describe event or problem, type of reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMAGER II ANGIOGRAPHIC CATHETER
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
MDR Report Key11003786
Report Number2134265-2020-17440
Device Sequence Number3477615
Product Code DQO
UDI-Device Identifier08714729355144
UDI-Public08714729355144
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K121694
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/18/2021
Device Model Number38270
Device Catalogue Number38270
Device Lot Number0000145249
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/26/2020
Supplement Date Received by Manufacturer12/02/2020
Initial Report FDA Received Date12/14/2020
Supplement Report FDA Received Date12/17/2020
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/18/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
92484513-FA
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient Weight53
-
-