• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 20ML SYR LUER LOCK TIP RP; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 20ML SYR LUER LOCK TIP RP; SYRINGE, PISTON Back to Search Results
Model Number 8881520657
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.  if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted. as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event or Problem Description
The customer reported that when they were mixing the drugs a small piece of plastic was noticed inside of the syringe.There was no patient harm reported.
 
Additional Manufacturer Narrative
H3 evaluation summary: the device history record (dhr) cannot be reviewed because the lot number is unknown.The manufacturing site received one syringe sample in an opened package.A complete investigation was performed.Visual inspection to the quality inspection standard (qis) was conducted.A large plastic string was identified on the inside diameter of the syringe barrel.The plastic string was loose inside the syringe barrel and was visible on the face and side of the rubber tip and extended between two of the ribs of the plunger rod.The most probable root cause for the plastic string is determined to have been introduced into the syringe during the assembly process.There is a potential for plastic strings to originate during the transport of the molded components, including the syringe plunger, from the molding machines to the assembly process.Blow over tubes are cleaned and inspected at prescribed intervals.Visual inspections are performed throughout the length of each production run and in the event issues such as particulate are identified, corrective actions are performed to eliminate the source.Per procedure, complaint trends are evaluated during the monthly corrective and preventative action (capa) meeting to determine if a capa is warranted.At this time, there is not enough information and a capa will not be initiated.However, these conditions will be communicated to the appropriate manufacturing and quality assurance personnel through a quality alert.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
20ML SYR LUER LOCK TIP RP
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
COVIDIEN
1222 sherwood rd
norfolk NE 68701
MDR Report Key11003859
Report Number1915484-2020-01199
Device Sequence Number14339802
Product Code FMF
UDI-Device Identifier20884521012155
UDI-Public20884521012155
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8881520657
Device Catalogue Number8881520657
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 11/24/2020
Supplement Date Received by Manufacturer11/24/2020
Initial Report FDA Received Date12/14/2020
Supplement Report FDA Received Date03/08/2021
Patient Sequence Number1
-
-