• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JJGC S.A. FACILITY ACQUA IMPL TI 2.9X14; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JJGC S.A. FACILITY ACQUA IMPL TI 2.9X14; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Model Number 140.739
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Unspecified Infection (1930); Fibrosis (3167)
Date of Event 10/29/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.  the loss of an endosseous dental implant after successful osseointegration and restoration is a known inherent risk of the procedure.Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia.Absence of a recurrent peri-implant infection with suppuration.Absence of implant mobility.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported early failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2020 in fdi 31.On (b)(6) 2020, loss of osseointegration was verified.Patient presented with poor oral hygiene.The device was forwarded to the manufacturer.At the event the patient experienced: peri-implantitis and mobility.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FACILITY ACQUA IMPL TI 2.9X14
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key11006239
Report Number3008261720-2020-06275
Device Sequence Number753480
Product Code DZE
UDI-Device Identifier07899878017059
UDI-Public07899878017059
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K150199
Number of Events Summarized1
Summary Report (Y/N)Y
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 12/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/25/2021
Device Model Number140.739
Device Catalogue Number140.739
Device Lot NumberXM991
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2020
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/14/2020
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-