| Model Number |
FGS-1100 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Event occurred within the year of 2020.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a synthes employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, that the patient¿s fibulink syndesmosis repair kit broke post-op.On (b)(6) 2020, the patient underwent a revision procedure as the suture bridge of the fibulink syndesmosis repair kit had broken and was no longer intact with the fibula link.Also, the head of the fibula tensioning cap had sheared off and was no longer part of the tensioning cap-fibula link junction.Initially, the patient was implanted with fibulink syndesmosis repair kit on (b)(6) 2020, with no reported injury or fall and did not know the device failed until clinical follow-up.The implant was removed without complications.The patient claims to have been compliant and non-weight bearing.This report involves one (1) fibulink® syndesmosis repair kit/ti.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: investigation summary.Customer quality investigation: the complaint device was not received for investigation.The following investigation is based on the image(s) provided.The image was reviewed, and the complaint condition could be confirmed, the suture can be noticed separated from the link which could have resulted in the revision surgery.Since the device was not returned, a dimensional inspection and a functional test were not able to be performed.A manufacturing record evaluation could not be performed as the lot number could not be determined from the image provided.A definitive assignable root cause could not be determined based on the provided information during the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history lot: no dhr as the lot # is unknown.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|