| Catalog Number |
H938743 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received and is currently awaiting evaluation completion.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a 5000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag leaked from the middle port (also reported as "between the spike port tube and the spike port bonding area").The leak was discovered during setup and preparation.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added.H4: the lot was manufactured from august 03, 2018 - august 04, 2018.H10: the device was received for evaluation.Unaided visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed which revealed a leak between the spike port tube and the spike port bonding area.The reported condition was verified.The cause of the condition was not determined; however, the most likely cause was due to inadequate or lack of cyclohexanone applied to the spike port cap tube when it was inserted to the spike port during the manufacturing process causing an incorrect bonding.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|