DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Battery Problem (2885)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/20/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted when additional information is provided.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) revealed a "no battery pack" error message.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
|
|
Event Description
|
It was reported that during routine check, the cardiosave intra-aortic balloon pump (iabp) revealed a "no battery pack" error message.Specifically, it was reported that the unit's screen showed an "x" in battery 2.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields: e1 (initial reporter, event site email), h6 (patient codes, evaluation method codes, result codes, conclusion codes).A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and found that battery 2 was missing from the pump which resulted in the reported issue.The fse noted there to be no failure of the pump, itself.The fse replaced the battery that was missing from the pump, and then performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|