• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Battery Problem (2885)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 11/20/2020
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted when additional information is provided.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) revealed a "no battery pack" error message.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
 
Event Description
It was reported that during routine check, the cardiosave intra-aortic balloon pump (iabp) revealed a "no battery pack" error message.Specifically, it was reported that the unit's screen showed an "x" in battery 2.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Updated fields: e1 (initial reporter, event site email), h6 (patient codes, evaluation method codes, result codes, conclusion codes).A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and found that battery 2 was missing from the pump which resulted in the reported issue.The fse noted there to be no failure of the pump, itself.The fse replaced the battery that was missing from the pump, and then performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key11008019
MDR Text Key224602241
Report Number2249723-2020-02149
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-