• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, STRYKERFLOW ELECTROCAUTERY PROBE, L TIP, 5MM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, STRYKERFLOW ELECTROCAUTERY PROBE, L TIP, 5MM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 0250070443
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Codes Burn(s) (1757); Burn, Thermal (2530)
Date of Event 04/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the patient was burned during procedure.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: unintentional tissue burns probable root cause: generator power malfunction; material/design error; conductive irrigant used; manufacturing/assembly error; user error.The reported failure mode will be monitored for future reoccurrence.Manufacturing date is unknown.H3 other text: 81.
 
Event or Problem Description
It was reported that the patient was burned during procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, STRYKERFLOW ELECTROCAUTERY PROBE, L TIP, 5MM
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11008650
Report Number0002936485-2020-00516
Device Sequence Number3813760
Product Code GCJ
UDI-Device Identifier07613327051872
UDI-Public07613327051872
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K934094
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0250070443
Device Catalogue Number0250070443
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/17/2020
Supplement Date Received by Manufacturer11/17/2020
Initial Report FDA Received Date12/15/2020
Supplement Report FDA Received Date02/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Weight147
-
-