| Model Number |
C2A12 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event unit is anticipated to return to applied medical for evaluation.A follow-up report will be provided upon the completion of investigation.
|
| |
|
Event or Problem Description
|
|
Name of procedure being performed: vnotes hysterectomy.Description transcribed from telephone call with applied medical representative 23nov20: polyurethane sheath has partially separated from the placement ring during vnotes procedure.The damaged alexis was removed, and a new product was opened - the procedure was completed using the alexis from the second package but the original gelport was used.No consequences or injury for patient.The tear occurred from bottom of sheath at the ring, the tear was approximately 2 inches in diameter.The product is in theatres 2 in quarantine.Patient status: no consequences or injury to the patient.Type of intervention: change of device: alexis component.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Upon visual inspection, engineering confirmed that the sheath had separated from the inner ring at the seal line.The inner ring of the event unit was also unflipped.Based on the condition of the returned unit, it is likely the reported event occurred when the event unit was pulled with a large force, which caused the inner ring to unflip and the sheath to separate from the inner ring.This event is not reportable as it is unlikely to cause or contribute to death or serious injury.This report is a follow-up report for medwatch # 01681.
|
| |
|
Event or Problem Description
|
|
Name of procedure being performed: vnotes hysterectomy.Description transcribed from telephone call with applied medical representative 23nov20: polyurethane sheath has partially separated from the placement ring during vnotes procedure.The damaged alexis was removed, and a new product was opened - the procedure was completed using the alexis from the second package but the original gelport was used.No consequences or injury for patient.The tear occurred from bottom of sheath at the ring, the tear was approximately 2 inches in diameter.The product is in theatres 2 in quarantine.Patient status: no consequences or injury to the patient.Type of intervention: change of device - alexis component.
|
| |
|