Model Number 71453211
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during tka surgery, after installing the legion ps high flex xlpe size 3-4 9mm insert, it was unstable in combination with the baseplate.Finally a same size insert was used to complete the surgery.Further details are unknown.
 
Event or Problem Description
It was reported that, during tka surgery, after installing the legion ps high flex xlpe size 3-4 9mm insert, it was unstable in combination with the baseplate.Finally a same size insert was used to complete the surgery.With a delay of less than 30 minutes.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device has signs of damage from attempted use.The device was manufactured in 2019.A dimensional inspection was attempted but the device was too damaged from the attempted insertion to obtain accurate measurements.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/sizing issue or a procedural error.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Case-2020-00030831-1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN PS HIGH FLEX XLPE SZ 3-4 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11010717
Report Number1020279-2020-07442
Device Sequence Number19731442
Product Code JWH
UDI-Device Identifier00885556035252
UDI-Public00885556035252
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71453211
Device Catalogue Number71453211
Device Lot Number19HM21340
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Initial Date Received by Manufacturer 11/26/2020
Supplement Date Received by Manufacturer12/09/2020
01/15/2021
Initial Report FDA Received Date12/15/2020
Supplement Report FDA Received Date12/16/2020
01/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
-
-