Catalog Number SPR1510W
Device Problem Burst Container or Vessel (1074)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/02/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation summary: device returned for evaluation.The balloon folds were open on the returned device.The device failed negative prep.Upon inflation the device failed to maintain pressure and liquid leaked from the balloon.Upon inspection a longitudinal burst was visible from the distal cone to the mid-working length.No issues noted with inner member at the glue port or the bifurcate.A 0.014" wire front and back-loaded with no issues.No other deformation was noted.If information is provided in the future, a supplemental report will be issued.
 
Event Description
During a procedure an attempt was made to use one sprinter otw ptca balloon catheter to treat a lesion.Negative prep was not performed.The lesion was not pre-dilated.The device did not pass through a previously deployed stent.It was reported that inflation difficulties occurred during balloon inflation and the balloon does not keep pressure when inflated to 2atm.The patient was reported to be alive with no injuries.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINTER OTW
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
MDR Report Key11010943
MDR Text Key221542231
Report Number9612164-2020-04925
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
PMA/PMN Number
P790017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/15/2022
Device Catalogue NumberSPR1510W
Device Lot Number220391422
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2020
Initial Date Manufacturer Received 12/10/2020
Initial Date FDA Received12/15/2020
Supplement Dates Manufacturer Received12/17/2020
Supplement Dates FDA Received12/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-