• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE TMZF HIP STEM #4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE TMZF HIP STEM #4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6020-0435
Medical Device Problem Codes Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Inflammation (1932); Pain (1994); Metal Related Pathology (4530)
Date of Event 01/31/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.Device not returned.
 
Event or Problem Description
It's alleged by the attorney, through the filing of a legal claim, that the patient underwent a right total hip arthroplasty in 2007 where he was implanted with a stryker hip system.It is further alleged that the patient began to experience pain and clicking with movement.X-rays performed on (b)(6) 2019 revealed severe trunnion failure.Blood work also revealed elevated levels of cobalt.This resulted in the patient to undergo right hip revision surgery on (b)(6) 2020.
 
Event or Problem Description
It's alleged by the attorney, through the filing of a legal claim, that the patient underwent a right total hip arthroplasty in 2007 where he was implanted with a stryker hip system.It is further alleged that the patient began to experience pain and clicking with movement.X-rays performed on (b)(6) 2019 revealed severe trunnion failure.Blood work also revealed elevated levels of cobalt.This resulted in the patient to undergo right hip revision surgery on (b)(6) 2020.
 
Additional Manufacturer Narrative
An event regarding fretting and abnormal ion level involving an accolade stem was reported.The event was not confirmed for abnormal ion level but fretting was confirmed based on medical review.Method & results: -device evaluation and results: visual inspection , material analysis, functional and dimensional inspections could not be performed as the device was not returned.Clinical review: a review of the provided medical records and x-rays by a clinical consultant indicated: no clinical or pmh, no primary tha op report or date of surgery, no imaging studies, no examination of explanted components or surgical pathology report.Based upon the revision operative report, trunnion failure appears to be confirmed as noted in the event description, but insufficient data is presented to create a medical report for this case -product history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.-complaint history review: review indicated there have been no other similar events for the reported lot.Conclusion: an event regarding fretting and abnormal ion level involving an accolade stem was reported.The event was not confirmed for abnormal ion level but fretting was confirmed based on medical review.The exact cause of the event could not be determined because insufficient medical information was provided.Further information such as device return, more medical documentation, patient history, demographics are needed.If additional information are received, this investigation will be reopened and re-evaluated.The reported accolade stem was mated with a lfit v40 cocr head.The cocr head has been identified to be within scope of an nc capa.However, the reported "abnormal ion level' is not within scope of the capa.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE TMZF HIP STEM #4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key11011211
Report Number0002249697-2020-02674
Device Sequence Number753287
Product Code LPH
UDI-Device Identifier04546540510563
UDI-Public04546540510563
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2020
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2012
Device Model Number6020-0435
Device Catalogue Number6020-0435
Device Lot Number23401702
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/18/2020
Supplement Date Received by Manufacturer12/18/2020
Initial Report FDA Received Date12/15/2020
Supplement Report FDA Received Date01/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age73 YR
Patient Weight259
-
-