• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH MODIFIED TRINKLE DRILLING ATTACH F/BPL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH MODIFIED TRINKLE DRILLING ATTACH F/BPL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 530.793
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The actual device has been returned, and is currently pending evaluation.Once the evaluation has been completed, a supplemental medwatch report will be sent accordingly.  udi: (b)(4).
 
Event or Problem Description
It was reported that the adapter of the attachment device was coming apart.It was not reported if the event occurred during a surgical procedure.It was not reported if there was a delay to a planned procedure.It was not reported if there was a spare device available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the attachment device had an unknown liquid on the ball bearings, the blue label ring was worn, and the coupling was loose.It was further determined that the device failed pretest for check function.Therefore, the reported condition that the device was coming apart was confirmed.The assignable root cause of this condition was determined to be traced to maintenance, which is user error/misuse/abuse.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODIFIED TRINKLE DRILLING ATTACH F/BPL
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key11011884
Report Number8030965-2020-09649
Device Sequence Number3813358
Product Code HWE
UDI-Device Identifier07611819802643
UDI-Public7611819802643
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number530.793
Device Lot Number8842303
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2020
Supplement Date Received by Manufacturer01/18/2021
Initial Report FDA Received Date12/15/2020
Supplement Report FDA Received Date01/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-