| Model Number |
206967 |
| Medical Device Problem Codes |
Failure to Calibrate (2440); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Hip ee is part of a recall.Case type / application: tha.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported issue ¿ hip ee is part of a recall.Case type / application: (b)(4).Product inspection ¿ the failure is confirmed as it is under the scope of nc hip end effectors (p/n 206967) failed characterization values.Also the serial number (b)(6) lies in the medical device recall list.Product history review - review of the device history records indicate 46 devices were manufactured under p/n 206967 lot 19330219 and 39 were accepted (including (b)(6) into final stock on 26 mar 2020.Review of qt revealed the reason for rejection was not related to the failure in this investigation.Complaint history review ¿ a review of complaints in catsweb and trackwise related to p/n 206967 lot 19330219 shows 00 additional complaints related to the failure in this investigation.Nc/capa history review - a review of stryker¿s nc/capa database indicated there have been an nc and capa associated with the product and failure mode reported in this event.Conclusion ¿ the event was confirmed because the serial number (b)(6) lies in the medical device recall list corresponding to nc.If additional information is received, then the complaint will be reopened.Per preventive maintenance is where an action occurs that identifies device deterioration which may compromise function.Under pm conditions no patient was involved, and no actual or potential patient harm existed for the alleged event.No additional investigation or specific actions are required.
|
| |
|
Event or Problem Description
|
|
Hip ee is part of a recall.Case type / application: (b)(4).
|
| |
|