• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012445-12
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device was discarded and not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation determined the reported balloon rupture appeared to be related to operational circumstances of the procedure.Based on the reported information, it is likely that during inflation, the balloon became damaged or compromised against the previously implanted stent resulting in the reported balloon rupture.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that on (b)(6) 2020, a percutaneous coronary intervention was performed on the left circumflex (cx) coronary artery, 90% stenosed lesion.An unspecified stent was implanted and a 2.0x12mm nc trek balloon dilatation catheter advanced without resistance for post-dilatation.During the initial inflation at 12 atmospheres, the balloon ruptured.The nc trek was removed without reported issues and another device, same size was used in replacement.There were no adverse patient effects and there was no clinically significant delay.No additional information was provided regarding this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11013011
Report Number2024168-2020-10530
Device Sequence Number4562882
Product Code LOX
UDI-Device Identifier08717648151644
UDI-Public08717648151644
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Model Number1012445-12
Device Catalogue Number1012445-12
Device Lot Number80929G1
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/27/2020
Initial Report FDA Received Date12/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-