| Model Number |
1012445-12 |
| Medical Device Problem Codes |
Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device was discarded and not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation determined the reported balloon rupture appeared to be related to operational circumstances of the procedure.Based on the reported information, it is likely that during inflation, the balloon became damaged or compromised against the previously implanted stent resulting in the reported balloon rupture.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2020, a percutaneous coronary intervention was performed on the left circumflex (cx) coronary artery, 90% stenosed lesion.An unspecified stent was implanted and a 2.0x12mm nc trek balloon dilatation catheter advanced without resistance for post-dilatation.During the initial inflation at 12 atmospheres, the balloon ruptured.The nc trek was removed without reported issues and another device, same size was used in replacement.There were no adverse patient effects and there was no clinically significant delay.No additional information was provided regarding this issue.
|
| |
|