| Model Number |
39200 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that stent damage occurred.The target lesion was located in anastomosis of fistula.A 6x60x75 epic vascular stent was selected for use.However, the distal stent struts where the markers lie are slightly peeled away from the catheter prior to placement into the sheath.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an epic self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that the stent is partially deployed 1mm from the distal end of the middle sheath.There is a kink to the sheath 58.5cm from the nosecone.Microscopic examination revealed that the sheath is flattened at the distal end.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.The target lesion was located in anastomosis of fistula.A 6x60x75 epic vascular stent was selected for use.However, the distal stent struts where the markers lie are slightly peeled away from the catheter prior to placement into the sheath.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|