Model Number 39200
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/01/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in anastomosis of fistula.A 6x60x75 epic vascular stent was selected for use.However, the distal stent struts where the markers lie are slightly peeled away from the catheter prior to placement into the sheath.The procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an epic self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that the stent is partially deployed 1mm from the distal end of the middle sheath.There is a kink to the sheath 58.5cm from the nosecone.Microscopic examination revealed that the sheath is flattened at the distal end.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in anastomosis of fistula.A 6x60x75 epic vascular stent was selected for use.However, the distal stent struts where the markers lie are slightly peeled away from the catheter prior to placement into the sheath.The procedure was completed with another of the same device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11013105
Report Number2134265-2020-17820
Device Sequence Number3944734
Product Code NIO
UDI-Device Identifier08714729805335
UDI-Public08714729805335
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P110035
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/27/2025
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0025933746
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2020
Initial Date Received by Manufacturer 12/08/2020
Supplement Date Received by Manufacturer01/14/2021
Initial Report FDA Received Date12/15/2020
Supplement Report FDA Received Date01/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-