The results of the investigation are inconclusive since the device was not returned for analysis. based on the information received, the cause of the reported incident could not be conclusively determined.However, the healthcare professional stated the patient's death was due to the patient's condition, and not a malfunction of the abbott product.
|
The patient passed away due to ventricular fibrillation (vf).The transmission on merlin.Net was reviewed showing eight non-sustained vf/ventricular tachycardia (vt) episodes.The egm showed an irregular rhythm with ventricular complexes with varying amplitudes consistent with a dying heart.Due to the autosense feature, intermittent undersensing was observed.The vf episode was detected, anti-tachycardia pacing was delivered; however, charging for defibrillation was aborted due to return to sinus.The undersensing present may have led to premature return to sinus.The healthcare professional did not allege a malfunction of the icd rather the event occurred due to the patient's condition.
|