ST. JUDE MEDICAL, INC. UNIFY ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3361-40C
Device Problem Under-Sensing (1661)
Patient Problems Arrhythmia (1721); Death (1802)
Event Date 12/08/2020
Event Type  Death  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.  based on the information received, the cause of the reported incident could not be conclusively determined.However, the healthcare professional stated the patient's death was due to the patient's condition, and not a malfunction of the abbott product.
 
Event Description
The patient passed away due to ventricular fibrillation (vf).The transmission on merlin.Net was reviewed showing eight non-sustained vf/ventricular tachycardia (vt) episodes.The egm showed an irregular rhythm with ventricular complexes with varying amplitudes consistent with a dying heart.Due to the autosense feature, intermittent undersensing was observed.The vf episode was detected, anti-tachycardia pacing was delivered; however, charging for defibrillation was aborted due to return to sinus.The undersensing present may have led to premature return to sinus.The healthcare professional did not allege a malfunction of the icd rather the event occurred due to the patient's condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11013641
MDR Text Key221706364
Report Number2938836-2020-09672
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Model NumberCD3361-40C
Device Lot NumberS000057485
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/05/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
-
-