| Model Number |
7015 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 95% stenosed, 21mm x 3.0mm target lesion was located in the severely calcified left coronary artery.A 3.0mm x 15mm quantum maverick balloon catheter was advanced for dilatation but the balloon burst.The procedure was completed with another of the same device.There were no complications reported and the patient is stable.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 95% stenosed, 21mm x 3.0mm target lesion was located in the severely calcified left coronary artery.A 3.0mm x 15mm quantum maverick balloon catheter was advanced for dilatation but the balloon burst.The procedure was completed with another of the same device.There were no complications reported and the patient is stable.It was further reported that the balloon was inflated once and ruptured at 15 atmospheres for 20 seconds.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast in the inflation lumen and balloon.The balloon was loosely folded.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There were no issues detected.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 95% stenosed, 21mm x 3.0mm target lesion was located in the severely calcified left coronary artery.A 3.0mm x 15mm quantum maverick balloon catheter was advanced for dilatation but the balloon burst.The procedure was completed with another of the same device.There were no complications reported and the patient is stable.It was further reported that the balloon was inflated once and ruptured at 15 atmospheres for 20 seconds.
|
| |
|