Model Number VCP375H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: could you please confirm if the suture separate from the needle or did the needle broke from the swage part? the needle broke from the swage part.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6)2020 and suture was used.During the procedure, the suture got detached from the needle.There were no adverse patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
(b)(4).Date sent to the fda: 2/1/2021.Additional h3 investigation summary: a failed suture needle was submitted for fractographic evaluation.A fracture was observed in the body of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was examined for this evaluation.A microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage and macroscopic plastic deformation observed coincidental to the fracture provide additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.The evidence of this examination indicates that the breakage occurred from mechanical deformation, indents and scratches, due to gripping, bending and overstressing of the needle.  there is no evidence of any material flaw that would cause premature failure.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ VIO 27IN 2-0 S/A UR-5
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key11015164
Report Number2210968-2020-09949
Device Sequence Number11661283
Product Code GAM
UDI-Device Identifier10705031052871
UDI-Public10705031052871
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model NumberVCP375H
Device Catalogue NumberVCP375H
Device Lot NumberQBMRMZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2021
Initial Date Received by Manufacturer 11/27/2020
Supplement Date Received by Manufacturer01/11/2021
Initial Report FDA Received Date12/15/2020
Supplement Report FDA Received Date02/01/2021
Patient Sequence Number1
-
-