| Model Number |
G407211 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Perforation (2513)
|
| Date of Event |
12/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported cardiac tamponade could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
During an atrial fibrillation ablation procedure, at the end a cardiac tamponade was noted via echocardiogram.A brk needle was used without issues during the procedure, although the transseptal puncture was noted to be difficult.The patient was noted to be stable and no pericardial drain was needed.The procedure was still completed.The perforation was believed to have occurred at the beginning of the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Per the ifu, cardiac perforation is a known risk during the use of this device.
|
| |
|