Model Number G407211
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 12/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported cardiac tamponade could not be conclusively determined.
 
Event or Problem Description
During an atrial fibrillation ablation procedure, at the end a cardiac tamponade was noted via echocardiogram.A brk needle was used without issues during the procedure, although the transseptal puncture was noted to be difficult.The patient was noted to be stable and no pericardial drain was needed.The procedure was still completed.The perforation was believed to have occurred at the beginning of the procedure.
 
Additional Manufacturer Narrative
Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK XS TRANSSEPTAL NEEDLE, 98 CM LENGTH
Common Device Name
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key11015263
Report Number3008452825-2020-00700
Device Sequence Number2651512
Product Code DRC
UDI-Device Identifier05414734205153
UDI-Public05414734205153
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K072278
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model NumberG407211
Device Catalogue NumberG407211
Device Lot Number7662652
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/16/2020
Supplement Date Received by Manufacturer12/16/2020
Initial Report FDA Received Date12/15/2020
Supplement Report FDA Received Date12/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-