Manufacturer's investigation conclusion: the report of elevated power consumption and rotor noise was confirmed via the submitted log files.A root cause for the power elevations and rotor noise could not be conclusively determined through this investigation.The submitted system controller event log 10 transient measurements of elevated power were captured on (b)(6) 2020 between 09:16 and 09:18.The power then returned to typical values for the remainder of the log file.The pump operated at the set speed for the duration of the log file.The system appeared to function as intended.The submitted lvad event log file captured rot_noise flags on (b)(6) 2020 between 09:16 and 09:18; aligning with the elevated power measurements captured within the controller event log file.The noise decreased to typical values at the end of the log.9 low flow measurements below 2.5 lpm were captured.Based on complaint history and similarly reported events, the data captured in the log files is potentially consistent with a deposition or thrombus formation passing through the system; however, a specific cause for the low flow, rotor instability, and rotor noise cannot be conclusively be determined through this evaluation as the pump was not returned.The relevant sections of the device history records for (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The heartmate 3 lvas ifu and the heartmate 3 lvas patient handbook are currently available.This ifu lists thromboemolism as a "potential risk and adverse event," as well as a "potential late postimplant complication" that may be associated with the use of the heartmate 3 left ventricular assist system.The patient care and management section of the ifu outlines indications of thrombus and how to respond to such events.This section also discusses anticoagulation, including recommended inr values.Section 1 of the ifu also addresses all pump parameters, including pump flow.Section 4, ¿system monitor¿, describes the pump flow display and the hazard alarms.This ifu states that the low flow hazard alarm will be triggered when pump flow is less than 2.5 lpm and explains that changes in patient conditions can result in low flow.Section 7, "alarms and troubleshooting", describes all alarm conditions, including the low flow hazard alarm, as well as the appropriate actions associated with them.No further information was provided.The manufacturer is closing this event.
|