Model Number 4720
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the anchor broke during the procedure.It was not able to be confirmed that all pieces were removed.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: anchor broken.Probable root cause: design.Inadequate raw material specification (needle or depth indicator).Needle tube thickness too thin or too thick.Suture does not have adequate working length.Inadequate handle assembly design.Inadequate raw material.Depth indicator thickness too thin or too thick process.Depth indicator not manufactured to specification.Needle or implant body not manufactured to specification.Dimple process on needle not manufactured to specification.Inadequate design of needle/depth indicator to create interference to retain/remove depth indicator.Stiffener not manufactured to specification application.User intentionally bends/forces instrument.Incision too small.Inadequate or improper visualization.User not familiar with device use.User does not use sled or other technique to prevent catching on soft tissue.The reported failure mode will be monitored for future reoccurrence.Manufacturing date is unknown.
 
Event or Problem Description
It was reported that the anchor broke during the procedure.It was not able to be confirmed that all pieces were removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIR MENISCUS SYSTEM, CURVED
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11016381
Report Number0002936485-2020-00532
Device Sequence Number14576733
Product Code GAT
UDI-Device Identifier07613327375404
UDI-Public07613327375404
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153087
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4720
Device Catalogue Number4720
Device Lot Number7002820
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/25/2020
Supplement Date Received by Manufacturer11/25/2020
Initial Report FDA Received Date12/16/2020
Supplement Report FDA Received Date01/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-