RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Temperature Problem (3022); Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
Malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020, date of report: 15dec2020.
|
|
Event Description
|
The customer reported cbit high temp blower alarm while in use.
The customer contacted product support and reported 1122 error in the significant event logs.
The customer advised air inlet is clean.
The customer advised cooling fan is working.
Product support advised the customer of the suspected blower assembly.
The customer reported that the unit was in use on a patient, but there was no patient harm reported.
The customer responded by email to product support and reported the issue resolved after replacing the blower assembly.
Patient information was requested from the customer, but no response was received.
|
|